Basic Troubleshooting Method For Thermal Shock Test Chamber

Aug. 09, 2018

1. In the Thermal Shock Test chamber, if the temperature change does not reach the test temperature value, the electrical system can be inspected and the faults can be eliminated one by one. If the temperature rises very slowly, it is necessary to check the air circulation system to see if the regulating baffle of the wind cycle is normal. On the contrary, check whether the motor of the wind cycle is running normally. If the temperature is too high, you need to set the PID setting parameters. If the temperature rises directly and the temperature is over-protected, then the controller fails and the control instrument must be replaced.

2. When the High And Low Temperature Test Chamber suddenly fails during the test operation, the corresponding fault display prompt appears on the control instrument and there is an audible alarm prompt. The operator of the thermal shock test chamber can quickly check which type of fault belongs to the fault diagnosis chapter in the operation of the equipment, and then the professional can quickly eliminate the fault to ensure the normal operation of the test. Other environmental testing equipment will have other phenomena in use, and it will be specific phenomena, specific analysis and elimination. The environmental test equipment should also be regularly maintained, the condenser of the refrigeration system should be cleaned regularly, the moving parts should be lubricated according to the instructions, the electrical control system should be regularly maintained and checked, etc. These tasks are essential.